SPRT–NTC Calibration Logging

Parallel data logging for thermistor (NTC) calibration against an SPRT reference. Two serial instruments are read simultaneously by a small command-line logger:

  • Isotech MicroK precision resistance bridge — reads the SPRT (reference thermometer).
  • SchwaRTech/AWI Temperature head — reads the NTCs (and optional test channels).

Both streams are timestamped (datetime) on every line, so the two data files can be matched up afterwards purely by their wall-clock timestamps.

Two ways to run:

  • calibration_log.pylegacy / passive. Logs both instruments; the bath temperature is set manually. It contains the whole logging machinery (both serial workers, the token-accurate NTC readout, the meta file) — calibration_auto.py imports it and only adds the bath control on top, so both tools write byte-identical data files.
  • calibration_auto.pyautomated. Additionally drives the Isotech Libra 785 bath through a list of temperature plateaus (via bisynch.py, EI-Bisynch over RS232) while logging continuously, ends each plateau when the SPRT reference stops drifting, and shows the live SPRT temperature next to the bath setpoint/PV so you can see at a glance where things stand.

Requirements

  • macOS (serial ports are addressed as /dev/cu.*)
  • Python 3
  • pyserial
python3 -m pip install -r requirements.txt

If import serial fails, make sure it is pyserial (python3 -m pip install pyserial) and not the unrelated PyPI package also called serial.

Hardware / wiring

Instrument Typical USB adapter Baud
MicroK bridge (SPRT) FTDI FT232R (single) 9600
SchwaRTech/AWI Temperature head FTDI FT2232H (dual, ch0) 19200

The exact /dev/cu.usbserial-* names can change between sessions — the script lets you pick the port interactively at runtime (see below). If you are unsure which device is on which port, tools/port_detect.py lists them and can probe them read-only (--list, --map config/param_combined.txt, --probe) — run it with the calibration stopped, one program per port.


Usage

Edit config/param_combined.txt first (see format below), then run:

python3 calibration_log.py                 # uses config/param_combined.txt
python3 calibration_log.py --param x.txt    # different config file
python3 calibration_log.py --exp Testlauf   # override the experiment name

At startup the script:

  1. lists the detected serial ports and asks you to pick which one is the MicroK and which is the SchwaRTech/AWI Temperature head (press Enter to accept the pre-selection from the config);
  2. asks for a free-text description of the calibration (until Enter) — this is stored in the _meta.txt file;
  3. starts both logging threads.

Stop with Ctrl-C — both threads shut down cleanly and close their ports. Robust for long, unattended runs (e.g. inside tmux/nohup over SSH).

Live output

Both raw values and the converted temperatures are shown on screen (the files keep the raw values only — the display conversion never touches them):

[MicroK] Port open : /dev/cu.usbserial-A922BJHF
[TempHead] Head is awake.
[MicroK] #1 Ch2: 0.234568   ->  -19.2540 C
[TempHead] 1: 5812827 | … || 4540193 | …   ->  N94=-46.661 N96=-39.454
...

The MicroK line adds the SPRT temperature (sprt.py); the head line adds the raw NTC temperature per node for every configured NTC channelNTC1, NTC2 and TestSB (which is wired as a second NTC) — via ntc.py. The NTC conversion uses the universal mean lab-S4 calibration (see below) and is for orientation only — the file contents are unchanged, so the R calibration pipeline is unaffected.

A node whose raw counts exceed 10_000_000 (open input) is reported as !! Nodes not connected: N91 instead of a temperature.

Output files can be copied at any time while logging — every line is flush()ed to disk immediately, so a copy always contains all completed lines.


Configuration — config/param_combined.txt

Minimal key: value format (order does not matter, # starts a comment):

experiment: Snowmelt_Retest5Sensors     # goes into the output file names

microk_channels: 1;2                     # Reference ; SPRT1 [ ; SPRT2 ]
microk_port: usbserial-A922BJHF          # optional port pre-selection

ntc_readout: NTC1;NTC2;TestSB            # from: TempADC, NTC1, NTC2, TestSB, TestN, GND, PRESSURE
ntc_groups: 5;60                         # legacy field, no longer chunks nodes (see below)
ntc_nodes: 90;91;92;93;94                # sensor node IDs
ntc_port: usbserial-FT3GCNKB0            # optional port pre-selection
Key Meaning
experiment Experiment name, used in output file names
microk_channels Bridge channels: Reference; SPRT1 (add ;SPRT2 for two SPRTs)
microk_port Optional MicroK port hint (pre-selection only)
ntc_readout Which sensor channels to log
ntc_groups Legacy — parsed for the meta file only; nodes are no longer chunked
ntc_nodes Sensor node IDs
ntc_port Optional Temperature-head port hint (pre-selection only)

Output — three files per run, in ./data/Output/ (four with calibration_auto.py)

<experiment>_<YYYYMMDD-HHMMSS>_microk.txt   # MicroK / SPRT data
<experiment>_<YYYYMMDD-HHMMSS>_ntc.txt      # NTC logger data
<experiment>_<YYYYMMDD-HHMMSS>_meta.txt     # resolved settings + verbatim copy of the param file

All three share the same timestamp, so one run = one matching set. The _meta.txt records the actually-chosen ports and all parameters, so the measurement settings are always recoverable later.

How the NTC readout works

The SchwaRTech/AWI temperature head is a multiplexed system: each physical sensor module is a node (ntc_nodes, e.g. 90;91;92;93;94). You tell the head which nodes to read with a single NODES 90 91 92 … command; it then streams one reading line per sweep containing exactly those nodes, and it keeps that node array in its own memory — across sweeps and even across restarts.

Because of that, the logger reads every node as one continuous array: it issues NODES once at startup and then just streams and logs, never re-issuing the command. (Earlier versions chunked the nodes into groups and round-robined between them, re-issuing NODES for each — but every command restarts the head, so with a short array that produced nothing but repeated header lines. Reading all nodes as one array removes the problem entirely.)

ntc_readout picks which channels each node outputs (NTC1, NTC2, TestSB, …); DateTime/TempADC are per-line, the rest are per-node. ntc_groups is a legacy field: it is still parsed (and echoed into the meta file) but no longer chunks the nodes — all nodes always go into the one array.

In the _ntc.txt file the header is written once at the start, followed by a continuous stream of data rows (all tagged Group1; for backward compatibility):

Group1; SecondsElapsed; DateTimePC; N94_NTC1 | N94_NTC2 | N94_TestSB || N96_NTC1 | N96_NTC2 | N96_TestSB
Group1; 9.19;  2026-07-02 15:59:58.58; 5812827 | 5732217 | 5648185 || 5540969 | 5471459 | 5400559
Group1; 13.77; 2026-07-02 16:00:03.15; 5310211 | 5253038 | 5192965 || 5113788 | 5066173 | 5013578
…
  • Within a data row, | separates channels within a node and || separates nodes. Every row fills all node columns — no blanks.
  • The header appears only once (not per cycle), so on a long run it scrolls out of a tail-read window; readers fall back to the known column layout from the config. New Node Array: / echo lines are recognised structurally and filtered out.
  • Rows that arrive incomplete (Ctrl-C, or the head stalling mid-line) are dropped rather than written, so every row in the file is full width.
  • If no data row arrives for 25 s (RESYNC_AFTER in calibration_log.py), the logger re-sends NODES once to nudge a genuinely stuck head — never more often, because every command restarts the head's sweep.
  • One array = one clean wide table, directly usable.

Per-value timestamps (TOFFMS). The head measures its sensors sequentially (number, |, next number, …), so a many-node line takes tens of seconds to over a minute to stream — and the bath oscillates ~10 mK on a ~2 min limit cycle. Tagging all values of a line with one timestamp would alias that oscillation into several mK of per-sensor error. The logger therefore reads the head byte-by-byte and records each value's true arrival time, appending a 5th ;-field to each data row:

Group1; 18.33; 2026-07-06 19:09:44.623; 861775 | … || … ; TOFFMS=0|763|1526|2290|…

TOFFMS holds one integer per value: milliseconds from the first value (whose time is the row's DateTimePC), in left-to-right order. So value k's absolute time is DateTimePC + TOFFMS[k] ms. Old 4-field readers ignore the field (the counts block is unchanged); the R pipeline uses it to pair each count with the SPRT interpolated to that exact instant, so the common-mode bath oscillation cancels instead of aliasing (see docs/DATA_FORMATS.md).


Automated bath-driven runs — calibration_auto.py

Same logging as the legacy tool, but it also steps the Isotech Libra 785 bath through a list of temperature plateaus.

How this bath is actually wired (verified on the unit). The Libra 785 has two controllers on one serial line, addressed separately and answered by the PC master one request at a time (never simultaneously — the master switches byte framing per request, exactly as Isotech's Cal NotePad does):

Controller Role Protocol Framing Address
Eurotherm 3504 main CONTROLLER — bath setpoint & PV EI-Bisynch 7E1 1
Isotech "OVER TEMPERATURE" safety limiter (read its temp only) Modbus RTU 8N1 2

So the bath is driven over EI-Bisynch (bisynch.py), not Modbus. Mnemonics in use: PV (measured), SL (writable setpoint), OP (% output), RR (ramp rate). bath.py (Modbus) still reads the over-temp limiter and remains available for other Isotech units whose controller is configured for Modbus. calibration_auto.py picks the path via the bath_protocol config key (default bisynch).

python3 -m pip install -r requirements.txt      # now also installs minimalmodbus
python3 calibration_auto.py --dry-run           # connect + read the bath, don't move it
python3 calibration_auto.py                      # full automated run (scrolling log)
python3 calibration_auto.py --dashboard          # fixed in-place overview panel

--dashboard replaces the scrolling log with a single panel that refreshes in place — plateau progress i/N, phase (settle/gate/dwell), bath PV/SP/OUT/ramp, the SPRT temperature, the live drift/sd gate metrics, and a live NTC table for all nodes. The loggers still write their files exactly as before; only the console view changes.

Quick bath-only usage — bath.py (no logging)

For just driving/reading the bath, bath.py is a self-contained CLI — no MicroK, no NTC, no config file. All modes read PV/SP/OUT once first as a smoke test.

python3 bath.py                                  # read PV / setpoint / output once
python3 bath.py --monitor                         # keep reading it out (Ctrl-C to stop)
python3 bath.py --set -40                          # set setpoint and exit
python3 bath.py --wait -40 --minutes 20            # set, wait until stable, hold 20 min
python3 bath.py --plateaus "-40;-20;0;20" --minutes 15   # run a ramp/plateau schedule
python3 bath.py --plateaus "0;25;50" --ramp-rate 5       # limited approach ramp (5 C/min)

Common flags: --encoding float|int1|int2|int3 (value scaling, default int1), --slave N, --tol, --window, --timeout. If PV/SP read back wrong, the encoding or baud/parity (in bath.py) is off — see the header notes.

Note: on the Libra 785, bath.py (Modbus) only reaches the over-temp limiter (slave 2). To drive the bath itself use the EI-Bisynch path — bisynch.py for one-off commands, or calibration_auto.py with bath_protocol: bisynch (default) for a full run.

Per plateau (calibration_auto.py): (optional ramp →) set setpoint → coarse settle (bath PV inside stability_tol for stability_window minutes) → soak until the plateau gate accepts → next. A fourth output file <experiment>_<time>_plateaus.txt records each plateau's setpoint, timestamps and gate metrics so the measurement window can be sliced out afterwards.

The plateau gate (gate_mode). A plateau is "done" not on a fixed clock but when the SPRT reference stops moving (gate_mode: drift, the default): accept the window once, over the trailing gate_window_min minutes, the fitted drift is < gate_drift_mk and the scatter is < gate_sd_mk. Slow creep (a few mK over half an hour) does no harm to the fit; a live dT/dt does — and the bath's own PV (10 mK resolution) cannot even see a 3 mK drift, so the gate reads the MicroK SPRT (sub-mK). plateau_minutes becomes the minimum soak before the gate may accept (long for the cold end, which creeps for hours); a hard cap (plateau_minutes + gate_max_extra_min) ends and flags any plateau whose reference never settles. gate_mode: fixed keeps the old fixed-time dwell. The accepted [t_dwell_start, t_dwell_end] in the plateau file is the trailing average window; the plateau-file header lists the exact columns (drift mode adds gate_ok; drift_mK; sd_mK; n).

The live status line shows bath PV/SP, the SPRT temperature (via sprt.py) and the raw NTC temperature per node for each NTC channel (NTC1/NTC2/TestSB, via ntc.py) so you see the whole picture at a plateau. Both conversions are display only — SPRT: the 2-point ratio calibration; NTC: the universal mean lab-S4 calibration (a 4-param Steinhart–Hart curve averaged over the 62 healthy GRIP sensors, ntc.py NTC_MEAN_COEF), which tracks a healthy sensor to ±0.03–0.05 °C and replaces the old β=3380 curve that read ~0.7 °C (0 °C) to ~3.6 °C (−40 °C) too warm. It is not a per-sensor fit — the authoritative per-sensor calibration stays in the R pipeline.

⚠️ Verify once against the real controller (see bath.py header): serial baud/parity, Modbus slave address, and the value encoding (bath_encoding: float vs int1/int2/int3). Run --dry-run first — if PV/SP read back as sensible numbers, the settings are right.

Everything lives in the single config/param_combined.txt, grouped into sections: (1) hardware — the three serial ports at the very top (they change when a USB-serial adapter is re-plugged); (2) experiment; (3) bath controller; (4) bath schedule & tuning. Bath-automation keys (ignored by the legacy tool):

Key Meaning
bath_port Bath-controller port hint (pre-selection)
bath_protocol bisynch (the 3504) | modbus (Series-2000 units)
bath_address Bisynch GID/UID address (the 3504 = 1) / Modbus slave address
bath_encoding Modbus only: float|int1|int2|int3 register scaling
plateaus 1–100 setpoints in °C, in run order (;-separated)
plateau_minutes drift mode: min soak before the gate; fixed mode: exact dwell. one → all, or per plateau
ramp_c_per_min Approach ramp; empty → max speed; one → all, or per plateau
stability_tol °C band around the setpoint counting as "arrived" (coarse)
stability_window minutes PV must stay in band before soaking
stability_timeout_per_10k max wait for arrival scales with step: minutes per 10 K
stability_timeout_min floor (minutes) for small/zero steps
gate_mode drift (SPRT dT/dt gate, default) | fixed (fixed-time dwell)
gate_drift_mk accept when |drift over the window| < this (mK), default 3
gate_sd_mk AND scatter about the fit line < this (mK), default 8
gate_window_min trailing window for drift+sd and the average (min), default 30
gate_max_extra_min soak cap = plateau_minutes + this (min), default 120
gate_poll_s how often the gate re-checks the SPRT log (s), default 30
gate_channel SPRT channel to gate on; default = first configured SPRT channel

If the coarse settle times out (step-scaled), the run soaks on the SPRT gate anyway; if the gate never passes within the soak cap, it measures the trailing window and flags the plateau. It never blocks the whole schedule.

A ready-to-run 24 h schedule is provided as config/param_24h.txt — an 11-point down-sweep with brackets (+5 … −35 in ~4 °C steps, 180 min soak at −35) plus 4 up-anchors (−25, −15, −5, +3) that re-visit interior temperatures later in time. The up-anchors make time orthogonal to temperature (common-mode drift is ~+1.5 mK/day), which keeps the fitted coefficients from bending:

python3 calibration_auto.py --param config/param_24h.txt --dashboard

Time estimate. At start (and, in --dashboard, continuously) the run prints a probable finish time and a latest finish time. "Latest" is a hard upper bound = Σ(arrival timeout + worst-case measure) per plateau — in drift mode the worst-case measure is the soak cap (plateau_minutes + gate_max_extra_min), so a plateau always ends; "probable" assumes each settles in ramp + a typical soak (≈ the empirical median).

"No communication with the instrument"

The port opened but the controller did not reply. Sweep the serial settings (read-only, changes nothing on the controller):

python3 bath.py --port /dev/cu.usbserial-XXXX --scan

It tries baud × parity × slave address and prints the first combination that answers, e.g. --baud 9600 --parity N --slave 1. Once found, pass those flags to the normal read-out.

If only the over-temp limiter answers Modbus (or nothing does), the main controller is on EI-Bisynch, not Modbus — that is the normal Libra 785 case. Find it read-only with the Bisynch scanner:

python3 bisynch.py --port /dev/cu.usbserial-XXXX --scan       # find framing + address
python3 bisynch.py --port ... --identify --addr 1             # map the mnemonics vs the panel
python3 bisynch.py --port ... --addr 1 --read PV              # read the measured temperature
python3 bisynch.py --port ... --addr 1 --write SL -15         # command a setpoint (changes the bath!)

--scan sweeps the Bisynch framings (7E1/7O1/8E1/8N1) × addresses and reports what answers; on this bath it is 7E1, address 1. --identify reads the common mnemonics so you can match them against the 3504 front panel before writing.


Repository layout

calibration_log.py      logger (MicroK + NTC head); also the library the auto tool imports
calibration_auto.py     automated run: same logging + bath plateau control
bisynch.py              EI-Bisynch control of the Eurotherm 3504 -- this drives the bath
bath.py                 Modbus RTU: the over-temp limiter here, the controller on Series-2000 units
gate.py                 SPRT dT/dt plateau gate (trailing drift/sd statistics)
sprt.py  ntc.py         ratio/counts -> temperature, live display only

config/                 parameter files -- param_combined.txt (default), param_24h.txt,
                        param_transient.txt
docs/                   HANDOVER.md (start here), DATA_FORMATS.md (output formats for R),
                        PID_TUNING.md (overshoot plan), TODO.md
tests/                  test_bath.py, test_bisynch.py -- offline, no hardware needed
tools/                  bath_sim.py (Modbus bath simulator), port_detect.py (which device
                        is on which port), requirements-sim.txt
data/Output/            measurement data written by a run (git-ignored)

Run the tools from the repository root -- the default parameter path (config/param_combined.txt) and the output directory (data/Output/) are relative to it. The test suites may be run from anywhere.


Authors

  • Nora Hirsch and Kathrin Brocker — original SPRT/NTC logging code.
  • The calibration team, with Claude Code — bath automation, plateau drift-gating, token-accurate logging, and the calibration workflow.